Personal tools

Sponsorname

Your company could advertise here

www.fla-consulting.eu

Reference:GBR Directive 93/42/EEC 11-Oct-2007 Article 9: Difference between revisions

From GxPlex

Jump to: navigation, search
m 1 revision imported
m 1 revision imported
 
(5 intermediate revisions by 4 users not shown)
Line 1: Line 1:
{{#set:
{{ReferenceData
|HasDocument=Document:GBR Directive 93/42/EEC 11-Oct-2007
|has_document=Document:GBR Directive 93/42/EEC 11-Oct-2007
|RefScope=MD
|module=1
|RefDocPart=Article 9
|unit=1.2.4
|RefLanguage=EN
|scope=MD
|RefTagsEN=
|doc_part=Article 9
|RefTagsOriginal=
|language=EN
|RefTagsFR=
|entry_by=Florian Adragna
}}
|contributors=
 
|tags=TAG-00053-Dev-las ++ TAG-00155-Med-Med ++ TAG-00198-Not-Not
{{:Document:GBR Directive 93/42/EEC 11-Oct-2007}}
|tags_display=[[Glossary:TAG-00053-Dev-las|Device class]], [[Glossary:TAG-00155-Med-Med|Medical device]], [[Glossary:TAG-00198-Not-Not|Notified body]]
 
|comment=Device classification rules (MDD)
{| class="gplx-infobox"
|content=Classification
|-
! colspan="2" class="gplx-infobox-header" | Reference Details
|-
! Scope
| MD
|-
! Document part
| Article 9
|-
! Language
| EN
|-
! Original entry by
| Florian Adragna
|-
! Contributors
|
|-
! Tags (EN)
|
|-
! Tags (original language)
|
|-
! Tags (FR)
|  
|-
! Comment
| Device classification rules (MDD)
|}
 
== Content ==
Classification
1. Devices shall be divided into Classes I, IIa, IIb and III. Classification shall be carried out in accordance with Annex IX.
1. Devices shall be divided into Classes I, IIa, IIb and III. Classification shall be carried out in accordance with Annex IX.
2. In the event of a dispute between the manufacturer and the notified body concerned, resulting from the application of the classification rules, the matter shall be referred for decision to the competent authority to which the notified body is subject.
2. In the event of a dispute between the manufacturer and the notified body concerned, resulting from the application of the classification rules, the matter shall be referred for decision to the competent authority to which the notified body is subject.
3. Where a Member State considers that the classification rules set out in Annex IX require adaptation in the light of technical progress and any information which becomes available under the information system provided for in Article 10, it may submit a duly substantiated request to the Commission and ask it to take the necessary measures for adaptation of classification rules. The measures designed to amend non-essential elements of this Directive relating to adaptation of classification rules shall be adopted in accordance with the regulatory procedure with scrutiny referred to in Article 7(3)
3. Where a Member State considers that the classification rules set out in Annex IX require adaptation in the light of technical progress and any information which becomes available under the information system provided for in Article 10, it may submit a duly substantiated request to the Commission and ask it to take the necessary measures for adaptation of classification rules. The measures designed to amend non-essential elements of this Directive relating to adaptation of classification rules shall be adopted in accordance with the regulatory procedure with scrutiny referred to in Article 7(3)
}}

Latest revision as of 13:49, 16 July 2026



Document information
Reference GBR Directive 93/42/EEC 11-Oct-2007
Validity area GBR
Scope(s) MD
Document name Directive 93/42/EEC
Version / Revision 11-Oct-2007
Status Current
Document type Legal act
Language(s) EN
Description COUNCIL DIRECTIVE 93/42/EEC - of 14 June 1993

concerning medical devices

Official source Official link
Restricted access No
Submitted by Florian Adragna
Contributors


Reference Details
Module Module 1 — Unit 1.2.4
Scope MD
Document part Article 9
Language EN
Original entry by Florian Adragna
Contributors
Tags Device class, Medical device, Notified body

Content

Classification 1. Devices shall be divided into Classes I, IIa, IIb and III. Classification shall be carried out in accordance with Annex IX. 2. In the event of a dispute between the manufacturer and the notified body concerned, resulting from the application of the classification rules, the matter shall be referred for decision to the competent authority to which the notified body is subject. 3. Where a Member State considers that the classification rules set out in Annex IX require adaptation in the light of technical progress and any information which becomes available under the information system provided for in Article 10, it may submit a duly substantiated request to the Commission and ask it to take the necessary measures for adaptation of classification rules. The measures designed to amend non-essential elements of this Directive relating to adaptation of classification rules shall be adopted in accordance with the regulatory procedure with scrutiny referred to in Article 7(3)