Reference:EU Regulation (EU) 2017/745 - Medical Device Regulation 20-Mar-2023 Article 87 Paragraph 3 – 5: Difference between revisions

From GxPlex

Jump to: navigation, search
gxplex>GxPlexBot
Import from GxPlex Excel
m 1 revision imported
 
(4 intermediate revisions by 4 users not shown)
Line 1: Line 1:
{{#set:
{{ReferenceData
|HasDocument=Document:EU Regulation (EU) 2017/745 - Medical Device Regulation 20-Mar-2023
|has_document=Document:EU Regulation (EU) 2017/745 - Medical Device Regulation 20-Mar-2023
|RefScope=MD
|module=5
|RefDocPart=Article 87
|unit=5.3.5
|scope=MD
|doc_part=Article 87
Paragraph 3 – 5
Paragraph 3 – 5
|RefLanguage=EN
|language=EN
|RefTagsEN=
|entry_by=Florian Adragna
|RefTagsOriginal=
|contributors=
|RefTagsFR=
|tags=TAG-00213-Ser-wie ++ TAG-00253-Rep-att ++ TAG-00066-Cau-enh ++ TAG-00181-Man-npr ++ TAG-00212-Inc-nis
}}
|tags_display=[[Glossary:TAG-00213-Ser-wie|Serious incident]], [[Glossary:TAG-00253-Rep-att|Reporting timeline]], [[Glossary:TAG-00066-Cau-enh|Causality]], [[Glossary:TAG-00181-Man-npr|Manufacturer]], [[Glossary:TAG-00212-Inc-nis|Incident]]
 
|comment=Reporting timelines for serious incidents
{{:Document:EU Regulation (EU) 2017/745 - Medical Device Regulation 20-Mar-2023}}
|content=3.  Manufacturers shall report any serious incident as referred to in point (a) of paragraph 1 immediately after they have established the causal relationship between that incident and their device or that such causal relationship is reasonably possible and not later than 15 days after they become aware of the incident.
 
{| class="wikitable gplx-infobox"
|-
! colspan="2" class="gplx-infobox-header" | Reference Details
|-
! Scope
| MD
|-
! Document part
| Article 87
Paragraph 3 – 5
|-
! Language
| EN
|-
! Original entry by
| Florian Adragna
|-
! Contributors
|
|-
! Tags (EN)
|
|-
! Tags (original language)
|
|-
! Tags (FR)
|  
|-
! Comment
| Reporting timelines for serious incidents
|}
 
== Content ==
3.  Manufacturers shall report any serious incident as referred to in point (a) of paragraph 1 immediately after they have established the causal relationship between that incident and their device or that such causal relationship is reasonably possible and not later than 15 days after they become aware of the incident.


4.  Notwithstanding paragraph 3, in the event of a serious public health threat the report referred to in paragraph 1 shall be provided immediately, and not later than 2 days after the manufacturer becomes aware of that threat.
4.  Notwithstanding paragraph 3, in the event of a serious public health threat the report referred to in paragraph 1 shall be provided immediately, and not later than 2 days after the manufacturer becomes aware of that threat.


5.  Notwithstanding paragraph 3, in the event of death or an unanticipated serious deterioration in a person's state of health the report shall be provided immediately after the manufacturer has established or as soon as it suspects a causal relationship between the device and the serious incident but not later than 10 days after the date on which the manufacturer becomes aware of the serious incident.
5.  Notwithstanding paragraph 3, in the event of death or an unanticipated serious deterioration in a person's state of health the report shall be provided immediately after the manufacturer has established or as soon as it suspects a causal relationship between the device and the serious incident but not later than 10 days after the date on which the manufacturer becomes aware of the serious incident.
}}

Latest revision as of 13:50, 16 July 2026



Document information
Reference EU Regulation (EU) 2017/745 - Medical Device Regulation 20-Mar-2023
Validity area EU
Scope(s) MD
Document name Regulation (EU) 2017/745 - Medical Device Regulation
Version / Revision 20-Mar-2023
Status Current
Document type Legal act
Language(s) EN
Description REGULATION (EU) 2017/745 OF THE EUROPEAN PARLIAMENT AND OF THE COUNCIL of 5 April 2017 on medical devices, amending Directive 2001/83/EC, Regulation (EC) No 178/2002 and Regulation (EC) No 1223/2009 and repealing Council Directives 90/385/EEC and 93/42/EEC
Official source Official link
Restricted access No
Submitted by Florian Adragna
Contributors


Reference Details
Module Module 5 — Unit 5.3.5
Scope MD
Document part Article 87

Paragraph 3 – 5

Language EN
Original entry by Florian Adragna
Contributors
Tags Serious incident, Reporting timeline, Causality, Manufacturer, Incident

Content

3. Manufacturers shall report any serious incident as referred to in point (a) of paragraph 1 immediately after they have established the causal relationship between that incident and their device or that such causal relationship is reasonably possible and not later than 15 days after they become aware of the incident.

4. Notwithstanding paragraph 3, in the event of a serious public health threat the report referred to in paragraph 1 shall be provided immediately, and not later than 2 days after the manufacturer becomes aware of that threat.

5. Notwithstanding paragraph 3, in the event of death or an unanticipated serious deterioration in a person's state of health the report shall be provided immediately after the manufacturer has established or as soon as it suspects a causal relationship between the device and the serious incident but not later than 10 days after the date on which the manufacturer becomes aware of the serious incident.