Reference:EU Regulation (EU) 2017/745 - Medical Device Regulation 20-Mar-2023 Annex XV Chapter I Section 2: Difference between revisions

From GxPlex

Jump to: navigation, search
gxplex>GxPlexBot
Import from GxPlex Excel
m 1 revision imported
 
(11 intermediate revisions by 4 users not shown)
Line 1: Line 1:
{{#set:
{{ReferenceData
|HasDocument=Document:EU Regulation (EU) 2017/745 - Medical Device Regulation 20-Mar-2023
|has_document=Document:EU Regulation (EU) 2017/745 - Medical Device Regulation 20-Mar-2023
|RefScope=MD
|module=1
|RefDocPart=Annex XV
|unit=1.2.8
|scope=MD
|doc_part=Annex XV
Chapter I
Chapter I
Section 2
Section 2
|RefLanguage=EN
|language=EN
|RefTagsEN=
|entry_by=Florian Adragna
|RefTagsOriginal=
|contributors=
|RefTagsFR=
|tags=TAG-00201-Cli-eSt ++ TAG-00203-Cli-lin ++ TAG-00338-Met-odi ++ TAG-00177-Ben-koB ++ TAG-00175-Per-ist
}}
|tags_display=[[Glossary:TAG-00201-Cli-eSt|Clinical investigation]], [[Glossary:TAG-00203-Cli-lin|Investigation plan]], [[Glossary:TAG-00338-Met-odi|Methodology]], [[Glossary:TAG-00177-Ben-koB|Benefit-risk]], [[Glossary:TAG-00175-Per-ist|Performance (device)]]
 
|comment=Clinical investigation methods and planning basis
== Document Information ==
|content=2.  Methods
{{:Document:EU Regulation (EU) 2017/745 - Medical Device Regulation 20-Mar-2023}}
 
== Reference Details ==
{| class="wikitable" style="width:100%;"
|-
! scope="row" style="width:200px;" | Scope
| MD
|-
! scope="row" style="width:200px;" | Document part
| Annex XV
Chapter I
Section 2
|-
! scope="row" style="width:200px;" | Language
| EN
|-
! scope="row" style="width:200px;" | Original entry by
| Florian Adragna
|-
! scope="row" style="width:200px;" | Contributors
|
|-
! scope="row" style="width:200px;" | Tags (EN)
|
|-
! scope="row" style="width:200px;" | Tags (original language)
|
|-
! scope="row" style="width:200px;" | Tags (FR)
|  
|-
! scope="row" | Comment
| Clinical investigation methods and planning basis
|}
 
== Text ==
2.  Methods


2.1.  Clinical investigations shall be performed on the basis of an appropriate plan of investigation reflecting the latest scientific and technical knowledge and defined in such a way as to confirm or refute the manufacturer's claims regarding the safety, performance and aspects relating to benefit-risk of devices as referred to in Article 62(1); the clinical investigations shall include an adequate number of observations to guarantee the scientific validity of the conclusions. The rationale for the design and chosen statistical methodology shall be presented as further described in Section 3.6 of Chapter II of this Annex.
2.1.  Clinical investigations shall be performed on the basis of an appropriate plan of investigation reflecting the latest scientific and technical knowledge and defined in such a way as to confirm or refute the manufacturer's claims regarding the safety, performance and aspects relating to benefit-risk of devices as referred to in Article 62(1); the clinical investigations shall include an adequate number of observations to guarantee the scientific validity of the conclusions. The rationale for the design and chosen statistical methodology shall be presented as further described in Section 3.6 of Chapter II of this Annex.
Line 57: Line 22:
[…]
[…]
2.8.  The clinical investigation report, signed by the investigator, shall contain a critical evaluation of all the data collected during the clinical investigation, and shall include any negative findings.
2.8.  The clinical investigation report, signed by the investigator, shall contain a critical evaluation of all the data collected during the clinical investigation, and shall include any negative findings.
}}

Latest revision as of 13:49, 16 July 2026



Document information
Reference EU Regulation (EU) 2017/745 - Medical Device Regulation 20-Mar-2023
Validity area EU
Scope(s) MD
Document name Regulation (EU) 2017/745 - Medical Device Regulation
Version / Revision 20-Mar-2023
Status Current
Document type Legal act
Language(s) EN
Description REGULATION (EU) 2017/745 OF THE EUROPEAN PARLIAMENT AND OF THE COUNCIL of 5 April 2017 on medical devices, amending Directive 2001/83/EC, Regulation (EC) No 178/2002 and Regulation (EC) No 1223/2009 and repealing Council Directives 90/385/EEC and 93/42/EEC
Official source Official link
Restricted access No
Submitted by Florian Adragna
Contributors


Reference Details
Module Module 1 — Unit 1.2.8
Scope MD
Document part Annex XV

Chapter I Section 2

Language EN
Original entry by Florian Adragna
Contributors
Tags Clinical investigation, Investigation plan, Methodology, Benefit-risk, Performance (device)

Content

2. Methods

2.1. Clinical investigations shall be performed on the basis of an appropriate plan of investigation reflecting the latest scientific and technical knowledge and defined in such a way as to confirm or refute the manufacturer's claims regarding the safety, performance and aspects relating to benefit-risk of devices as referred to in Article 62(1); the clinical investigations shall include an adequate number of observations to guarantee the scientific validity of the conclusions. The rationale for the design and chosen statistical methodology shall be presented as further described in Section 3.6 of Chapter II of this Annex. […] 2.4. Clinical investigations shall be performed in accordance with the clinical investigation plan by a sufficient number of intended users and in a clinical environment that is representative of the intended normal conditions of use of the device in the target patient population. Clinical investigations shall be in line with the clinical evaluation plan as referred to in Part A of Annex XIV. […] 2.6. The endpoints of the clinical investigation shall address the intended purpose, clinical benefits, performance and safety of the device. The endpoints shall be determined and assessed using scientifically valid methodologies. The primary endpoint shall be appropriate to the device and clinically relevant. […] 2.8. The clinical investigation report, signed by the investigator, shall contain a critical evaluation of all the data collected during the clinical investigation, and shall include any negative findings.