Reference:EU Regulation (EU) 2017/746 - in vitro diagnostic medical devices 20-Mar-2023 Article 58 Paragraph 5-002

From GxPlex

Revision as of 15:06, 29 June 2026 by gxplex>GxPlexBot (Import from GxPlex Excel)
Jump to: navigation, search



Document information
Reference EU Regulation (EU) 2017/746 - in vitro diagnostic medical devices 20-Mar-2023
Validity area EU
Scope(s) IVD
Document name Regulation (EU) 2017/746 - in vitro diagnostic medical devices
Version / Revision 20-Mar-2023
Status Current
Document type Legal act
Language(s) EN
Description Regulation (EU) 2017/746 of the European Parliament and of the Council of 5 April 2017 on in vitro diagnostic medical devices and repealing Directive 98/79/EC and Commission Decision 2010/227/EU
Official source Official link
Restricted access No
Submitted by Florian Adragna
Contributors


Reference Details
Scope IVD
Document part Article 58

Paragraph 5

Language EN
Original entry by Florian Adragna
Contributors
Tags (EN)
Tags (original language)
Tags (FR)
Comment Repeated entry: 1.2.7 ++ 2.3.2

Conditions required before conducting a performance study (repeated)

Content

5. A performance study as referred to in paragraph 1 may be conducted only where all of the following conditions are met:

(a) the performance study is the subject of an authorisation by the Member State(s) in which the performance study is to be conducted, in accordance with this Regulation, unless otherwise stated;

(b) an ethics committee, set up in accordance with national law, has not issued a negative opinion in relation to the performance study, which is valid for that entire Member State under its national law;

(c) the sponsor or its legal representative or a contact person pursuant to paragraph 4 is established in the Union;

(d) vulnerable populations and subjects are appropriately protected in accordance with Articles 59 to 64;

(e) the anticipated benefits to the subjects or to public health justify the foreseeable risks and inconveniences and compliance with this condition is constantly monitored;

(f) the subject or, where the subject is not able to give informed consent, his or her legally designated representative has given informed consent, in accordance with Article 59;

(g) the subject or, where the subject is not able to give informed consent, his or her legally designated representative, has been provided with the contact details of an entity where further information can be received in case of need;

(h) the rights of the subject to physical and mental integrity, to privacy and to the protection of the data concerning him or her in accordance with Directive 95/46/EC are safeguarded;

(i) the performance study has been designed to involve as little pain, discomfort, fear and any other foreseeable risk as possible for the subjects, and both the risk threshold and the degree of distress are specifically defined in the performance study plan and constantly monitored;

(j) the medical care provided to the subjects is the responsibility of an appropriately qualified medical doctor or, where appropriate, any other person entitled by national law to provide the relevant patient care under performance study conditions;

(k) no undue influence, including that of a financial nature, is exerted on the subject, or, where applicable, on his or her legally designated representatives, to participate in the performance study;

(l) where appropriate, biological safety testing reflecting the latest scientific knowledge or any other test deemed necessary in the light of the device's intended purpose has been conducted;

(m) in the case of clinical performance studies, the analytical performance has been demonstrated, taking into consideration the state of the art;

(n) in the case of interventional clinical performance studies, the analytical performance and scientific validity has been demonstrated, taking into consideration the state of the art. Where, for companion diagnostics, the scientific validity is not established, the scientific rationale for the use of the biomarker shall be provided;

(o) the technical safety of the device with regard to its use has been proven, taking into consideration the state of the art as well as provisions in the field of occupational safety and accident prevention;

(p) the requirements of Annex XIV are fulfilled.