Personal tools

Sponsorname

Your company could advertise here

www.fla-consulting.eu

Reference:EU Regulation (EU) 2017/746 - in vitro diagnostic medical devices 20-Mar-2023 Article 77 Paragraph 4 – 5

From GxPlex

Revision as of 15:29, 29 June 2026 by Gxplex admin (talk | contribs) (1 revision imported)
Jump to: navigation, search



Document information
Reference EU Regulation (EU) 2017/746 - in vitro diagnostic medical devices 20-Mar-2023
Validity area EU
Scope(s) IVD
Document name Regulation (EU) 2017/746 - in vitro diagnostic medical devices
Version / Revision 20-Mar-2023
Status Current
Document type Legal act
Language(s) EN
Description Regulation (EU) 2017/746 of the European Parliament and of the Council of 5 April 2017 on in vitro diagnostic medical devices and repealing Directive 98/79/EC and Commission Decision 2010/227/EU
Official source Official link
Restricted access No
Submitted by Florian Adragna
Contributors


Reference Details
Scope IVD
Document part Article 77

Paragraph 4 – 5

Language EN
Original entry by Florian Adragna
Contributors
Tags (EN)
Tags (original language)
Tags (FR)
Comment Sponsor reporting at end of performance study to all Member States

Content

4. If a study is conducted in more than one Member State, the sponsor shall notify all Member States in which that performance study was conducted of the end of the performance study in all Member States. That notification shall be made within 15 days of that end of the performance study.

5. Irrespective of the outcome of the performance study, within one year of the end of the performance study or within three months of the early termination or temporary halt, the sponsor shall submit to the Member States in which a performance study was conducted a performance study report as referred to in Section 2.3.3. of Part A of Annex XIII.

The performance study report shall be accompanied by a summary presented in terms that are easily understandable to the intended user. Both the report and summary shall be submitted by the sponsor by means of the electronic system referred to in Article 69.

Where, for scientific reasons, it is not possible to submit the performance study report within one year of the end of the study, it shall be submitted as soon as it is available. In such case, the clinical performance study plan referred to in Section 2.3.2. of Part A of Annex XIII shall specify when the results of the performance study are going to be available, together with a justification.