Reference:EU Regulation (EU) 2017/746 - in vitro diagnostic medical devices 20-Mar-2023 Article 82 Paragraph 3 – 5

From GxPlex

Revision as of 12:56, 16 July 2026 by GxPlexBot (talk | contribs) (Migrate to Template:ReferenceData for PageForms (GxPlexBot))
Jump to: navigation, search



Document information
Reference EU Regulation (EU) 2017/746 - in vitro diagnostic medical devices 20-Mar-2023
Validity area EU
Scope(s) IVD
Document name Regulation (EU) 2017/746 - in vitro diagnostic medical devices
Version / Revision 20-Mar-2023
Status Current
Document type Legal act
Language(s) EN
Description Regulation (EU) 2017/746 of the European Parliament and of the Council of 5 April 2017 on in vitro diagnostic medical devices and repealing Directive 98/79/EC and Commission Decision 2010/227/EU
Official source Official link
Restricted access No
Submitted by Florian Adragna
Contributors


Reference Details
Module Module 5 — Unit 5.3.5
Scope IVD
Document part Article 82

Paragraph 3 – 5

Language EN
Original entry by Florian Adragna
Contributors
Tags Serious incident, Field safety corrective action, Manufacturer, Incident, IVD

Content

Reporting of serious incidents and field safety corrective actions

3. Manufacturers shall report any serious incident as referred to in point (a) immediately after they have established a causal relationship between that incident and their device or that such causal relationship is reasonably possible, and not later than 15 days after they become aware of the incident.

4. Notwithstanding paragraph 3, in the event of a serious public health threat the report referred to in paragraph 1 shall be provided immediately, and not later than 2 days after the manufacturer becomes aware of that threat.

5. Notwithstanding paragraph 3, in the event of death or an unanticipated serious deterioration in a person's state of health the report shall be provided immediately after the manufacturer has established or as soon as it suspects a causal relationship between the device and the serious incident but not later than 10 days after the date on which the manufacturer becomes aware of the serious incident.