§ 812.46 Monitoring investigations […] (b) Unanticipated adverse device effects. (1) A sponsor shall immediately conduct an evaluation of any unanticipated adverse device effect. (2) A sponsor who determines that an unanticipated adverse device effect presents an unreasonable risk to subjects shall terminate all investigations or parts of investigations presenting that risk as soon as possible. Termination shall occur not later than 5 working days after the sponsor makes this determination and not later than 15 working days after the sponsor first received notice of the effect. [...]